Espec Temperature Chamber (Horizontal type) - PH(H)

ตู้อบ
Model: PH-102, PH-202, PH-302, PH-402, PHH-102, PHH-202, PHH-302, PHH-402.

สอบถามรายละเอียดเพิ่มเติมติดต่อ 085-550-5313, email: sale@esdscience.com

Features

  • High performance chamber
    A temperature-indication controller with an advance PID operation, and an originally developed chamber configuration provide unmatched oven performance. Temperature uniformity, temperature constancy, temperature heat-up rate, and temperature recovery time are performed with the upmost reliability.
  • Safety measures
    Triple safety mechanisms are employed for excessive overheating.
  • Wide model selection
    We provide a total of 16 ovens with combination of temperature range, capacity, and instrumentation
  • Energy Saver Duct (Option)
    Energy saving approximately 20% by heat recycling from exhaust through the dust to maintain temperature while damper opens.

Specification

Model

PH-102

PH-202

PH-302

PH-402

PHH-102

PHH-202

PHH-302

PHH-402

System

Forced hot-air circulation / ventilation system

Performance*1

Temperature range*2

((Ambient temp. +20) ℃) to +200℃

((Ambient temp. +20) ℃) to +300℃

Temperature fluctuation*2

±0.1℃ at + 100℃

±0.2℃ at + 200℃

±0.2℃ at + 100℃

±0.4℃ at + 200℃

±0.1℃ at + 100℃

±0.2℃ at + 200℃

                 + 300℃

±0.2℃ at + 100℃

±0.4℃ at + 200℃

±0.6℃ at + 300℃

Temperature uniformity*2

±0.5℃ at + 100℃

±1.5℃ at + 200℃

±1.0℃ at + 100℃

±2.0℃ at + 200℃

±0.5℃ at + 100℃

±1.5℃ at + 200℃

±2.5℃ at + 300℃

±1.0℃ at + 100℃

±2.0℃ at + 200℃

±3.0℃ at + 300℃

Temperature heat-up time

Ambient temp. to +200℃

Ambient temp. to +300℃

within 40 min.

within 60 min.

within 60 min.

within 70 min.

Construction

Exterior material

Cold rolled rust-proof steel plate, Melamine resin coating

Interior material

Stainless steel plate

Insulation material

Glass wool

Heater

Iron chrome strip wire heater

Air circulator

Stainless steel propeller fan

Damper

Circulation / Ventilation (manual switching)

Fittings

Power cable (approx. 2m from chamber), Specimen power supply control terminals

(relay contact is opened/stop during malfunction. Voltage capacity 250V AC 3A)

Inside dimensions W×H×D (in)

450×450×450

(17.7×17.7×17.7)

600×600×600

(23.6×23.6×23.6)

800×800×800

(31.5×31.5×31.5)

1000×1000×1000

(39.4×39.4×39.4)

450×450×450

(17.7×17.7×17.7)

600×600×600

(23.6×23.6×23.6)

800×800×800

(31.5×31.5×31.5)

1000×1000×1000

(39.4×39.4×39.4)

Outside dimensions *3 W×H×D (in)

1040×820×635

(41×32.3×25)

1190×970×785

(46.9×38.2×30.9)

1500×1210×1065

(59.1×47.6×41.9)

1730×1480×1275

(68.1×58.3×50.2)

1040×820×635

(41×32.3×25)

1190×970×785

(46.9×38.2×30.9)

1500×1210×1065

(59.1×47.6×41.9)

1730×1480×1275

(68.1×58.3×50.2)

Capacity (L)

91

216

512

100

91

216

512

1000

Weight (kg)

95

130

240

430

95

130

240

430

Allowable ambient conditions

Temperature: 0 to +40℃  Humidity: to 75%rh

Utility requirements

Power supply (Voltage fluctuation: ±10% of rated value)

200 / 220 / 230 / 240V AC

1 ⌀ 50/60Hz

200 / 220V AC

3 ⌀ 3W 50/60Hz,

380V AC 3 ⌀ 4W 50Hz

200 / 220 / 230 / 240V AC

1 ⌀ 50/60Hz

200 / 220V AC

3 ⌀ 3W 50/60Hz,

380V AC 3 ⌀ 4W 50Hz

Max. power consumption (kVA)

2.0

2.7

5.0

6.5

2.7

3.8

6.5

9.5

*1 Values assume circulatory operation with no specimens at an ambient temperature of +23℃±5.
*2 Conforms to Japan Testing Machinery standard K05:2000.
*3 Excluding protrusions.